Turbomolecular vacuum pumps NVT-450
Turbomolecular vacuum pumps NVT-450 are used for gases and gas mixtures (air) extraction from vacuum systems (with the exception of hydrocarbons, aggressive gases, vapors that can affect construction materials of the pump) and creation of deep and superdeep vacuum.
- surface physics and gas analysis, manufacturing of electronic devices;
- leak detection, semiconductors production;
- creation of oil-free initial vacuum for electrophysical extraction devices-ion, evaporation (getter) and cryogenic pumps, combined (getter-ion) pumps, exhaust carts, units etc.
The pumps are supplied in a moderately cold climate modification category 4 according to GOST 15150. Ion pumps are powered by special power supply units (further – BPM), not included in the delivery package of the pump. The pumps are supplied in a moderately cold climate modification category 4 according to GOST 15150. Turbomolecular pumps NVT-450 require a forevacuum system for their operation. Turbomolecular pumps are powered by special power supply units (further – BPT), not included in the pump delivery package.
NOTE: Power supply units BPT for turbomolecular the pumps are supplied by a separate order, together with the pump or separately. Key features of pumps are shown in the table below. Design and basic dimensions of counter flanges of turbomolecular vacuum pumps are shown in the Appendix. Basic specifications of BPT power supply units recommended for our pumps are shown in General specifications of turbomolecular vacuum pumps.
Specifications of turbomolecular vacuum pumps NVT-450
Parameter |
NVT-450 FR |
NVT-450 FO |
NVT-450 FG |
Pump speeds with air (without inlet protection screen) over the pressure range of 1*10-2 - 1*10-4 Pa, ± 20 %, not less than l/sec |
400 |
400 |
400 |
Ultimate residual pressure of the pump (at discharge pressure not more than 6 Pa), Pa |
1*10-5 |
6,5*10-6 |
6,5*10-6 |
Acceleration time, minutes |
10 |
10 |
10 |
Rotational speed, sec-1 |
410±8 |
410±8 |
410±8 |
General levels of vibroacceleration measured in the horizontal plane of the pump inlet flange, in the established mode, not more than, dB |
90 |
90 |
90 |
Cooling |
water-cooled |
water-cooled |
water-cooled |
Grease |
consistent |
consistent |
consistent |
Maximum temperature of the pump inlet flange preheating, °С |
45 |
45 |
45 |
Counter outlet (high-vacuum) flange of the pump, mm |
Mounting flange 160 RD 11 868.006 |
Toothed flange 160 OST 11 868.007 |
Dn-160 Flange GOST 26526 |
Counter exhaust (forevacuum) flange of the pump, mm |
Flange 40 OST 11.868.005 |
Flange 40 OST 11.868.005 |
Flange 40 OST 11.868.005 |
Pump dimensions, not more than, mm* |
250 х 260 х 300 |
280 х 280 х 290 |
250 х 260 х 280 |
Weight, not more than, kg* |
22 |
28 |
27 |
Recommended power supply unit |
BPT-1000 |
BPT-1000 |
BPT-1000 |
Maximum BPT power consumption, not more than, V*A |
500 |
500 |
500 |
Recommend forevacuum pump |
2NVR-5DM |
2NVR-5DM |
2NVR-5DM |
* dimensions and weight without shipping plugs